Modeling of FiO2 delivered by self-inflating ventilation device
Modelagem de FiO2 entregue por bolsa auto inflável
DOI:
https://doi.org/10.34117/bjdv8n11-003rKeywords:
fraction of inspired oxygen, FiO2 modeling, bag resuscitatorAbstract
During the COVID-19 pandemic, professionals in the medical field have been using bag resuscitators, also called bag valve masks, to provide respiratory support for patients for long periods. A relevant issue arising because of such use is the little information regarding the FiO2 delivered by these devices, leading to toxic effects of O2 and its waste. This paper presents a model that relates volume, respiratory frequency, and flow in an O2 line with the FiO2 delivered by bag resuscitators. The model was obtained based on a formal procedure and data from a 1500 ml Hsiner bag resuscitator collected in the literature, both presented in this paper. Then, we compared through Bland-Altman analysis the FiO2 values estimated by our model with the ones from data considering different bag resuscitators to evaluate the amplitude of the errors it can generate. The average error for Hsiner and Laerdal bag resuscitators was less than 0.035 (or 3.5%) and less than 0.063 (or 6.3%) for Ambu devices. The maximum error found for Ambu bag resuscitators was high. The obtained model presented satisfactory errors, compatible with the uncertainty levels associated with FiO2 measuring instruments and blenders, meaning the model can be used to estimate the delivered FiO2 accurately. Additionally, the procedure presented in this paper can be used to obtain specific models for bag resuscitators of other companies, equipping professionals with tools that can accurately estimate FiO2.
References
WORLD HEALTH ORGANIZATION. Coronavirus disease (covid-19) pandemic [Access in Feb 28 2022]. Available in: https://www.who.int/emergencies/diseases/novel-coronavirus-2019.
MATTHAY, M. A.; ALDRICH, J. M.; GOTTS, J. E. Treatment for severe acute respiratory distress syndrome from covid-19. The Lancet Respiratory Medicine, v.8, n. 5, p.433-434, 2020. Available in: https://doi.org/10.1016/S2213-2600(20)30127-2.
SHIM, E.; TARIQ, A.; CHOI, W.; LEE, Y.; CHOWELL, G. Transmission potential and severity of covid-19 in south korea. International Journal of Infectious Diseases, v. 93, p.339-344, 2020. Available in: https://doi.org/10.1016/j.ijid.2020.03.031.
WUJTEWICZ, M.; DYLCZYK-SOMMER, A; LOWICZ, A. A.; ZDANOWSKI, S.; PIWOWARCZYK, S.; OWCZUK, R. Covid-19 - what should anesthesiologists and intensivists know about it? Anaesthesiology Intensive Therapy, v. 52, n. 1, p.34-41, 2020. Available in: https://doi.org/10.5114/ait.2020.93756.
LOFORTE, A.; DAL CHECCO, E.; GLIOZZI, G.; BENEDETTO, M.; CAVALLI, G. G.; MARIANI, C.; PICCONE, G.; AGULLI, M.; PACINI, D.; BAIOCCHI, M. Venovenous extracorporeal membrane oxygenation support in covid-19 respiratory distress syndrome: initial experience. Asaio Journal, 2020. Available in: https://doi.org/10.1097/MAT.0000000000001198
GRASSELLI, G; TONETTI, T.; PROTTI, A.; LANGER, T.; GIRARDIS, M.; BELLANI, G.; LAEY, J.; CARRAELLO, J.; CARSANA, L.; RIZZUTO, C.; et al. Pathophysiology of covid-19-associated acute respiratory distress syndrome: a multicentre prospective observational study. The Lancet Respiratory Medicine, v. 8, n. 12, p.1201-1208, 2020. Available in: https://doi.org/10.1016/S2213-2600(20)30370-2.
PONS-ÒDENA, M.; VALLS, A.; GRIFOLS, J.; FARRÉ, R.; LASOSA, F. J. C.; RUBIN, B. K. Covid-19 and respiratory support devices. Paediatric Respiratory reviews, 2020. Available in: https://doi.org/10.1016/j.prrv.2020.06.015.
CHANG, J.; ACOSTA, A.; BENAVIDES, J.; REATEGUI, J.; ROJAS, C.; COOK, J.; NOLE, R.; GIAMPIETRI, L.; PÉREZ, S.; CASADO, F.; et al. Masi: A mechanical ventilator based on a manual resuscitator with telemedicine capabilities for patients with ards during the covid-19 crisis. HardwareX, p.e00187, 2021. Available in: https://doi.org/10.1016/j.ohx.2021.e00187.
GINO, B.; WANG, Z.; D'ENTREMONT, P.; RENOUF, T. S.; DUBROWSKI, A. Automated Inflating resuscitator (air): Design and development of a 3d-printed ventilator prototype and corresponding simulation scenario based on the management of a critical covid-19 patient. Cureus, v. 12, n. 7, 2020. Available in: https://doi.org/10.7759/cureus.9134.
KWON, A. H.; SLOCUM, A. H.; VARELMANN, D.; NABZDYK, C. G. S. Rapidly scalable mechanical ventilator for the covid-19 pandemic. Intensive care medicine, v. 46, n. 8, p.42-1644, 2020. Available in: https://doi.org/10.1007/s00134-020-06113-3.
URBINA, J.; MONKS, S. M.; OCHOA, L.; STUMP, R. F.; WICKER, R. B.; DANEK, C.; TORRES, V. I.; CRAWFORD, S. B. Development and evaluation of an automated manual resuscitator-based emergency ventilator-alternative. Cureus, v. 13, n. 3, 2021. Available in: https://doi.org/10.7759/cureus.13642.
KALYAEV, V.; SALIMON, A. I.; KORSUNSKY, A. M.; DENISOV, A. A. Fast mass-production of medical safety shields under covid-19 quarantine: Optimizing the use of university fabrication facilities and volunteer labor. International journal of environmental research and public health, v. 17, n. 10, p.3418, 2020. Available in: https://doi.org/10.3390/ijerph17103418.
DIMPEL, L. Self-inflating ventilation devices and FiO2. Anaesthesia, v. 64, n. 11, p.1268-1269, 2009. Available in: https://doi.org/10.1111/j.1365-2044.2009.06120.x.
GODOY, A. C. F. Avaliação em bancada do desempenho de sete diferentes marcas de ressuscitadores manuais auto-infláveis para adultos produzidos e/ou comercializados no Brasil. PhD thesis, Universidade Estadual de Campinas, Faculdade de Ciências Médicas, Campinas - SP, 2007.
SMITH, T. P.; MAGUIRE, D. Effects of tidal volume, ventilation rate and oxygen flow on delivered O2 during manual ventilation with a resuscitation bag. The anesthesiology annual meeting, 2012.
GODOY, A. C. F.; VIEIRA, R. J. Comparison of the O2 delivered by seven models of the self-inflating bag-mask system. Revista brasileira de anestesiologia, v. 59, n. 1, p.21-27, 2009. Available in: https://doi.org/10.1590/S0034-70942009000100004.
KIM, K. J.; NAM, Y. T.; KU, M. W.; CHON, S. S.; KOH, S. O. The changes in delivered oxygen fractions using laerdal resuscitator bag with corrugated tubes. Korean Journal of Anesthesiology, v. 38, n. 2, p.327-332, 2000. Available in: http://doi.org/10.4097/kjae.2000.38.2.327.
YOUNG, P.; PILCHER, J.; PATEL, M.; CAMERON, L.; BRAITHWAITE, I.; WEATHERALL, M.; BEASLEY R. Delivery of titrated oxygen via a self-inflating resuscitation bag. Resuscitation, v. 84, n. 3, p.391-394, 2013. Available in: https://doi.org/10.1016/j.resuscitation.2012.08.330.
ZAMORA I.; NEWHART, J. The influence of variable flows on o2 delivery via self-inflating resuscitation bags, 2018
NAM, S. H.; KIM, K. J.; NAM, Y. T.; SHIM, J. K. The changes in delivered oxygen fractions using laerdal resuscitator bag with different types of reservoir. Yonsei Medical Journal, v. 42, n. 2, p.242-246, 2001.
QUINTANA, S.; PÉREZ, J. P.; ALVAREZ, M.; VILA, J. S.; JARA, F.; NAVA, J. M. Maximum FiO2 in minimum time depending on the kind of resuscitation bag and oxygen flow. Intensive care medicine, v. 30, n. 1, p.155-158, 2004. Available in: https://doi.org/10.1007/s00134-003-2010-x.
GODOY, A. C. F.; VIEIRA, R. J.; VIEIRA-NETO, R. J.; et al. Oxygen out-flow delivered by manually operated self-inflating resuscitation bags in patients breathing spontaneously. Jornal brasileiro de Pneumologia, 2008. Available in: https://doi.org/10.1590/s1806-37132008000400005.
IRUJO, J. J. A.; VELASCO, J. M.; MORAL, P.; CARRILLO, B.; LÓPEZ-TIMONEDA, F. Delivered oxygen fraction during simulated cardiopulmonary resuscitation depending on the kind of resuscitation bag and oxygen flow. European Journal of Emergency Medicine, v. 19, n. 6, p.359-362, 2012. Available in: https://doi.org/10.1097/MEJ.0b013e32834dd360.
IRUJO, J. J. A. Findings in chalcogens reported from hospital clinic. Respiratory Therapeutics Week, p.263, 2012.
YOUNG, P. Findings in the area of respiratory research reported from university of otago. Respiratory Therapeutics Week, p.33, 2013.
SANZ-SANJOSÉ, E.; IRUJO, J. J. A.; MARTÍN, C. E. S.; PERRINO, C. G.; LÓPEZ-TIMONEDA, f. Mínimo flujo de oxígeno necesario para soporte vital durante la simulación de reanimación post parada cardiorrespiratoria. Revista Española de Anestesiología y Reanimación, v. 63, n. 5, p.261-266, 2016. Available in: https://doi.org/10.1016/j.redar.2015.08.002.
TUSMAN, G.; CAMPOS, M.; GOGNIAT, E. Covid-19: How to transform a noninvasive ventilation device in a critical care ventilator. Revista Española de Anestesiología y Reanimación (English Edition), v. 67, n. 7, p.367-373, 2020. Available in: https://doi.org/10.1016/j.redare.2020.05.008.
KULKARNI, A. P.; DIVATIA, J. V.; PATIL, V. P.; GEHDOO, R. P. Objective Anaesthesia Review: A Comprehensive Textbook for the Examinees. JP Medical Ltd, 2013.
CHAPRA, S. C.; CANALE, R. Numerical Methods for Engineers. McGraw-Hill, 5th edition, 2005. ISBN 0073101567.
BLAND, J. M.; ALTMAN, D. G. Measuring agreement in method comparison studies. Statistical methods in medical research, v. 8, n. 2, p.135-160, 1999. Available in: https://doi.org/10.1177/096228029900800204.
JAKETMP; TIRRELL, L. Pycompare version 1.5.1 [Access in Aug 28 2022]. Available in: jaketmp/pycompare.
PRECISIONMEDICAL. Precisionmedical: Air-oxygen blender (user manual), 2020.
MARKOVCHICK, V. J. Emergency Medicine Secrets E-Book. Elsevier Health Sciences, 2010.